72191202 (05/2012) 0.0 METTLER TOLEDO DIVA Service Manual PSC 8200 Magellan Scanner
METTLER TOLEDO DIVA Service Manual 4 Mettler-Toledo (ChangZhou) Scale & System Ltd. Legal Metrology Declaration of Conformity Konformitätser
METTLER TOLEDO DIVA Service Manual 5 Important notice for verified weighing instruments M M M M Weighing Instruments verified at the place of manufac
METTLER TOLEDO DIVA Service Manual 6 2 Installation 2.1 Precautions Before you install your DIVA scale, keep in mind the following factors, which m
METTLER TOLEDO DIVA Service Manual 7 For Weigh Only Version: The Weigh Only version comes with a tower mount. Assemble the Weight Only tower as is
METTLER TOLEDO DIVA Service Manual 8 2.5 Sealing After installation is complete, legal-for-trade applications require sealing the enclosure so the s
METTLER TOLEDO DIVA Service Manual 9 For your notes:
METTLER TOLEDO DIVA Service Manual 10 3 Set-up and Calibration There are two types of setup modes. The first is called “Service Mode” and it allo
METTLER TOLEDO DIVA Service Manual 11Press the Zero Key: Price Computing Version Weigh Only Version You can press the Tare key to
METTLER TOLEDO DIVA Service Manual 12 3.3 Set up Soft Switch and Default Table Group.Step Function Possible Selections Master Mode Access Group
METTLER TOLEDO DIVA Service Manual 13 Group.Step Function Possible Selections Master Mode AccessGroup 3 Press Zero to EnterPress Tare to go to Gro
METTLER TOLEDO DIVA Service Manual 14 3.4 Country Defaults PC WO WO WO WO WO WO WO WO WO WO WO WO WO WO WOFunctionGermany/ France USA Austria Spain
METTLER TOLEDO DIVA Service Manual 153.5 GeoCal Country Codes Country Geo-Value Austria 18 Belgium 21 Bulgaria 16 Croatia 18 Czechia 20 Denmark 23 E
METTLER TOLEDO DIVA Service Manual 16 3.6 GeoCal USA State Location Codes State Code Alabama Birmingham & North South of Birmingham 13 12 A
METTLER TOLEDO DIVA Service Manual 17
METTLER TOLEDO DIVA Service Manual 18 3.7 Calibration Note: These functions are directly related to the Weight & Measurement regulations in you
METTLER TOLEDO DIVA Service Manual 19Note: If the scale is unable to complete calibration it will display “E 20”. Cycle power to the scale and rest
METTLER TOLEDO DIVA Service Manual 20 4 Operating Instructions 4.1 Displays 72190532 72190533 72187467 72187468 7218
METTLER TOLEDO DIVA Service Manual 21 4.2 Keyboard Key Description Used to return the scale to gross zero. Tare key. 4.3 Cursors Cursor Descri
METTLER TOLEDO DIVA Service Manual 22 4.4 Operations 4.4.1 Weighing and Communication Place the item to be weighed on the platter, the weight will
METTLER TOLEDO DIVA Service Manual 235 Service and Maintenance This chapter provides information on servicing and maintaining the scale including:
INTRODUCTIONThis publication is provided solely as a guide for individuals who have purchased the METTLER TOLEDO Diva scale product. Information r
METTLER TOLEDO DIVA Service Manual 24 5.3 Flash Updating the Software The Diva’s software can be flash updated through the serial communications por
METTLER TOLEDO DIVA Service Manual 255.5 Parts and Descriptions General Exploded Diagram.
METTLER TOLEDO DIVA Service Manual 26 Note: Most parts are not available for individual replacement. This following chart and diagrams are onl
METTLER TOLEDO DIVA Service Manual 275.6 Replacement Parts Order Numbers Part Number Exploded ID Description Picture 72191106 45-53 SL DISP,W/LENS
METTLER TOLEDO DIVA Service Manual 28 Part Number Exploded ID Description Picture 71209398 Not Shown Mounting Screw (PSC, same as Dura) 7218576
METTLER TOLEDO DIVA Service Manual 296 Appendix 6.1 Protocols The numbers correspond to the number that will show up in the Soft Switch settings in Gr
METTLER TOLEDO DIVA Service Manual 30 6.1.1 NCI – ECR 6.1.1.1 OVERVIEW The serial communications protocol is used to interface Weigh-Tronix/NCI PO
METTLER TOLEDO DIVA Service Manual 31 Name: Zero the scale Command: Z<CR> Response: Scale is zeroed, returns scale status. (se
METTLER TOLEDO DIVA Service Manual 32 Weight Command: W This command causes the scale to return the displayed weight string in ECR format. The scal
METTLER TOLEDO DIVA Service Manual 33 Scale Status Command: S There will be at least two status bytes. If bit 6 of the second status byte is set then
Precautions WARNING ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THIS EQUIPMENT. EXERCISE CARE WHEN MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE
METTLER TOLEDO DIVA Service Manual 34 6.1.2 (Reserved)
METTLER TOLEDO DIVA Service Manual 356.1.3 8217 Mettler Toledo 6.1.3.1 Host Communications The 8217 scale acts as a peripheral device when connect
METTLER TOLEDO DIVA Service Manual 36 6.1.3.2 Command Descriptions ASCII Commands (* MUST BE UPPERCASE CHARACTERS ONLY.) *ASCII Command From Host S
METTLER TOLEDO DIVA Service Manual 376.1.3.3 Status Byte Format When the scale responds with STX?status byte CR , this indicates a status byte. The s
METTLER TOLEDO DIVA Service Manual 38 6.1.3.4 Confidence Test Status Byte Format When the scale responds with STX? status byte CR, after requesting
METTLER TOLEDO DIVA Service Manual 396.1.3.5 Tare Tare can be taken on an item two ways: Automatic or digital. REMOTE TARE The unknown weight of an o
METTLER TOLEDO DIVA Service Manual 40 6.1.4 8213 Mettler Toledo 6.1.4.1 Host Communications The 8217 scale acts as a peripheral device when conne
METTLER TOLEDO DIVA Service Manual 416.1.4.2 Command Descriptions ASCII Commands (* MUST BE UPPERCASE CHARACTERS ONLY.) *ASCII Command From Host Scal
METTLER TOLEDO DIVA Service Manual 42 6.1.4.3 Status Byte Format When the scale responds with STX?status byte CR , this indicates a status byte. Th
METTLER TOLEDO DIVA Service Manual 436.1.4.4 Confidence Test Status Byte Format When the scale responds with STX? status byte CR, after requesting a
DIVA Service Manual 3 Contents 1 INTRODUCTION ...
METTLER TOLEDO DIVA Service Manual 44 6.1.5 EPOS 1 6.1.5.1 Data Format 2400 Baud (select 2400, 4800 or 9600) 7 data bits (select 7 or 8) Even parit
METTLER TOLEDO DIVA Service Manual 45 6.1.5.3 Command Flow ECR Scale COMMENTSENQ (05h) ACK (06h) Data Available. Errors could be: CAN - repeat we
METTLER TOLEDO DIVA Service Manual 46 6.1.6 EPOS 2 6.1.6.1 Data Format 2400 Baud (select 2400, 4800 or 9600) 7 data bits (select 7 or 8) Even parit
METTLER TOLEDO DIVA Service Manual 47 6.1.6.4 Command Flow ECR Scale COMMENTSENQ (05h) ACK (06h) Data Available. Errors could be: CAN - repeat we
METTLER TOLEDO DIVA Service Manual 48 6.1.7 (Reserved)
METTLER TOLEDO DIVA Service Manual 496.1.8 Dialog 06 With free programmable POS-Systems it is for a third party on principle possible, to manipulate p
METTLER TOLEDO DIVA Service Manual 50 6.1.8.3 Command Descriptions Records for Communications from a POS system to the Scale Record 01: Transmitt
METTLER TOLEDO DIVA Service Manual 51 Records for Communications from the Scale to a POS system Record 02: Result of weighing (answer to recor
METTLER TOLEDO DIVA Service Manual 52 6.1.8.4 Description of the Dialog A dialogue will always be initiated by the POS-system. Description of wei
METTLER TOLEDO DIVA Service Manual 53 6.1.8.5 Flow Chart EOT STX SATZ 03 ETXSTXSATZ 04 ETXETXSTX SATZ 05 ETXSTX SATZ 10 ETXSTX SATZ 20 ETXENQSTX
METTLER TOLEDO DIVA Service Manual 54 6.1.9 Dialog 04 / 02 Note: To use Dialog 04, select this protocol but set the baud rate to 4800, for Dialog
METTLER TOLEDO DIVA Service Manual 55 6.1.9.3 Command Descriptions POS System Communications to a Scale Setting number 01 - Basic Price EOT STX 3
METTLER TOLEDO DIVA Service Manual 56 Scale Communications to a POS System Setting number 02: Weight, Price, and Total Price STX 30h 32h ESC
METTLER TOLEDO DIVA Service Manual 57 6.1.10 Reserved
METTLER TOLEDO DIVA Service Manual 58 6.1.11 ICL 6.1.11.1 Data Format 9600 Baud (select 2400, 4800 or 9600) 7 data bits (select 7 or 8) Even parit
METTLER TOLEDO DIVA Service Manual 59 6.1.11.4 Command Flow ECR Scale COMMENTSENQ (05h) ACK (06h) Data Available. Errors could be: CAN - repeat w
METTLER TOLEDO DIVA Service Manual 60 METTLER TOLEDO 1900 Polaris Parkway Columbus, Ohio 43420-2020 P/N: 72191204
METTLER TOLEDO DIVA Service Manual 11 Introduction The DIVA is designed to mount on the Metrologic Stratos Bi-Optic Scanners and communicate with ECR
METTLER TOLEDO DIVA Service Manual 2 1.3 Physical Dimensions The dimensions of the scale with tower is as follows:
METTLER TOLEDO DIVA Service Manual 3 1.4 Power Requirements Power is supplied via a Y-cable connected to the Scanner power supply. 1.5 Temperature
Comments to this Manuals