METTLER TOLEDO
DIVA Service Manual 35
Name: Zero the scale
Command: Z<CR>
Response: Scale is zeroed, returns scale status. (see note 4)
<LF>Shhhh<CR><ETX>
Optional Commands
Name: Request high-resolution weight
Command: H<CR>
Response: Returns decimal wt in 10x with units plus scale status. (see notes 2
and 3)
<LF>xxx.xxxUU<CR><LF>Shhhh<CR><ETX>
Returns lb-oz wt in10x with units plus scale status.
<LF>xLB<SP>xx.xxOZ<CR><LF>Shhhh<CR><ETX>
Scale status only if wt < 0, initial zero error, in motion or out of capacity.
<LF>Shhhh<CR><ETX>
Name: Change units of measure
Command: U<CR>
Response: Changes units of measure, returns new units and scale status.
<LF>uu<CR><LF>Shhhh<CR><ETX>
Name: Request metrology raw counts
Command: M<CR>
Response: Returns normalized raw counts and scale status.
<LF>xxxxxxMM<CR><LF>Shhhh<CR><ETX>
Name: unrecognized command
Command: all others
Response: Unrecognized command
<LF>?<CR><ETX>
NOTES:
1) Weight field is always six characters (5 for weight, 1 for decimal point), regardless
of display size.
2) Leading zeroes are not suppressed.
3) High resolution weight field is seven characters (6 for weight, 1 for decimal point),
regardless of display size.
4) If scale is outside zero range or in motion the ‘Z’ command is ignored and scale
status is returned without zero status flag set, i.e., scale not at zero.